Cardiologia Clínica e Experimental

Cardiologia Clínica e Experimental
Acesso livre

ISSN: 2155-9880

Abstrato

A Case of Thrombus Aspiration Catheter Trapped in Coronary Artery Due To Rupture of its Main Shaft and Twisting Over the Wire

Cagdas Akgullu, Ufuk Eryılmaz, Hasan Gungor, Mucahit Avcil, Bekir Daglı, Imran Kurt Omurlu and Cemil Zencir

We report a novel complication that arose while treating a 54 year old man with acute inferior ST elevation segment myocardial infarction. It was decided to perform thrombus aspiration prior to balloon angioplasty. However, after aggressive maneuvers to cross the tight lesion in order to suck the distal thrombus, a novel complication was encountered. The Thrombus Aspiration Catheter (TAC) got stuck in the coronary artery. When the whole apparatus was pulled out, it was realized that the guidewire cut the main shaft of TAC and catheter got twisted over the wire. To the best of our knowledge, this is the first report of a complication of this type.

Isenção de responsabilidade: Este resumo foi traduzido com recurso a ferramentas de inteligência artificial e ainda não foi revisto ou verificado.